3T Heater-Cooler Manufacturer Issues Updated Infection Prevention Info



DOCUMENTS
  • Letter


WASHINGTON, D.C. — The maker of the 3T Heater-Cooler System, LivaNova, has issued a letter informing health care providers of ways to reduce potential cardiac surgery infection associated with the device, according to an alert from the Food and Drug Administration.

In the letter, LivaNova (formerly Sorin Group Deutschland GmbH), has “provided updated instructions to monitor the concentration of hydrogen peroxide in the water circuit to verify that sufficient concentration of hydrogen peroxide is present to limit microbial growth, and to adjust the concentration of hydrogen peroxide if it drops below 100 ppm,” the FDA explained in an Oct. 22 …






UPCOMING CONFERENCES




HarrisMartin's MDL Conference: Video Game Addiction and the Latest Mass Tort Updates

May 29, 2024 - Salt Lake City, UT
The Grand America Hotel

MORE DETAILS



HarrisMartin's Mass Tort Settlements Conference - Sponsored by Milestone

June 27, 2024 - Buffalo, NY
The Richardson Hotel Buffalo

MORE DETAILS